What Is AQL 1.5? A Complete Guide to Disposable Glove Quality Testing and Medical Glove Standards
Fujian Jixiang E-commerce Co., Ltd

What Is AQL 1.5? (Direct Answer)
AQL (Acceptable Quality Level) 1.5 is the international standard for medical-grade examination gloves. It means that in a production batch, no more than 1.5% of gloves are allowed to have pinhole defects (holes that would compromise the barrier between the wearer and contaminants). AQL 1.5 is defined in EN 455-1 and ASTM D3578 / D6319 and is the minimum requirement for gloves labeled as "medical-grade" or "examination grade."
In simple terms: if a factory produces 10,000 gloves and tests a statistical sample, no more than 1.5% of the tested gloves can fail the water leak test. If more than 1.5% fail, the entire batch is rejected.

Why AQL 1.5 Matters
The AQL rating directly determines whether a glove can be used in medical settings where barrier protection against bloodborne pathogens is critical. Key implications:
• AQL 1.5 → Medical examination gloves, surgical gloves (minimum)
• AQL 2.5 → Industrial-grade gloves (some food-grade applications)
• AQL 4.0 → General-purpose / utility gloves (non-medical)
How AQL Testing Works: The Water Leak Test
AQL testing follows a standardized procedure defined in EN 455-1:
-
Sampling: A statistical sample is drawn from the production batch based on ISO 2859-1 sampling tables.
-
Water Leak Test: Each glove in the sample is filled with 1,000 mL of water, suspended, and observed for 2–5 minutes.
-
Inspection: Any visible water leakage = a failed glove (pinhole defect).
-
Decision: If the number of failed gloves exceeds the AQL 1.5 threshold for the sample size, the entire batch is rejected.
| Batch Size | Sample Size | Max Allowed Failures (AQL 1.5) |
|---|
| 2,801 – 10,000 | 200 | 7 |
| 10,001 – 35,000 | 315 | 10 |
| 35,001 – 150,000 | 500 | 14 |
| 150,001 – 500,000 | 800 | 21 |
Source: ISO 2859-1 Sampling Tables, Normal Inspection Level II
AQL 1.5 vs AQL 2.5 vs AQL 4.0
| AQL Level | Defect Rate Allowed | Typical Use | Standards |
|---|
| AQL 0.65 | ≤ 0.65% | Surgical gloves | EN 455-1 (surgical) |
| AQL 1.5 | ≤ 1.5% | Medical examination gloves | EN 455-1, ASTM D3578 |
| AQL 2.5 | ≤ 2.5% | Industrial / food-grade gloves | EN 374, ASTM D6319 |
| AQL 4.0 | ≤ 4.0% | General utility gloves | EN 420 |
Which Standards Require AQL 1.5?
| Standard | Scope | AQL Requirement |
|---|
| EN 455-1 | Medical gloves for single use | AQL 1.5 (examination), AQL 0.65 (surgical) |
| ASTM D3578 | Latex examination gloves | AQL 1.5 |
| ASTM D6319 | Nitrile examination gloves | AQL 1.5 |
| ASTM D5250 | PVC examination gloves | AQL 1.5 |
| GSO 4872 | GCC medical glove standard | AQL 1.5 (aligned with EN 455) |
Note: If a glove does not state its AQL on the packaging, it is likely an industrial-grade glove (AQL 2.5 or higher) and should NOT be used for medical applications.
How to Verify AQL 1.5 as a Buyer
-
Request the Certificate of Analysis (COA) — This document should show AQL test results for each batch.
-
Check the packaging — Medical-grade gloves must display the AQL level and relevant standard (e.g., "AQL 1.5 per EN 455-1").
-
Ask for third-party test reports — Reputable suppliers will provide SGS, TUV, or Intertek test reports verifying AQL compliance.
-
Conduct your own sampling — For large orders, consider independent pre-shipment inspection (PSI) with AQL testing.
FAQ
Q: What does AQL stand for?
A: AQL stands for Acceptable Quality Level. It is a statistical sampling method defined in ISO 2859-1 that determines the maximum number of defective units allowed in a production batch before the entire batch is rejected.
Q: Is AQL 1.5 good enough for medical gloves?
A: Yes. AQL 1.5 is the international standard for medical examination gloves under EN 455-1 and ASTM standards. Surgical gloves require a stricter AQL 0.65. Any glove with AQL > 1.5 should not be used for medical purposes.
Q: How can I tell if a glove is truly AQL 1.5?
A: Check the glove box labeling — it should clearly state "AQL 1.5" and reference the applicable standard (EN 455-1 or ASTM D3578/D6319). Request the Certificate of Analysis (COA) and third-party test reports from the supplier.
Q: Does AQL 1.5 apply to both nitrile and latex gloves?
A: Yes. AQL 1.5 applies to all examination glove materials — nitrile (ASTM D6319), latex (ASTM D3578), and PVC/vinyl (ASTM D5250). The AQL requirement is material-independent; what matters is the intended use (medical vs. non-medical).

Contact us at www.jxgloves.com for samples and documentation.